TM 10-5410-221-14/1
Continuity must exist
observing the following:
between wire ends. Replace wire if continuity does not
(1) Open circuit breaker panel cover (125, fig. 6-7,
exist between wire ends.
sheet 5 of 5).
f.
Repair.
Repair 400 Hertz wiring harness
(2) Remove circuit breaker panel per paragraph 6-
21b(2).
assembly to return to a serviceable condition as follows.
(3) Remove terminal board cover (35, fig. 6-7, sheet
(1) Inspect individual wires for cuts and abrasions of
2 of 5) and remove nut (29), lockwasher (30) and washer
insulation. Minor cuts and abrasions are acceptable and
(31) from terminal 5on terminal board (38) for the
shall be repaired by wrapping damaged area with vinyl
harness assembly(5, fig. 6-7, sheet 1 of 5) and terminal 4
tape or other suitable electrical wiring tape. Replace any
for the harness assembly (6) and remove the electrical
individual wires which are cut through or otherwise
leads from the terminal boards.
damaged.
(4) Remove nuts (1), washers (2) and screws (3) and
(2) Inspect tie straps (10, fig. 6-26) for tightness and
remove harness assemblies (5 and 6) from distribution
fraying. Replace any loose or frayed tie straps.
g. Installation. Install 400 Hertz wiring harness in
box. Remove cover (4).
(5) Remove screws to disconnect the harness
reverse order of removal using figure 6-7 as a guide, and
electrical leads from circuit breakers (101, fig. 6-7, sheet
observing the following:
5 of 5) for the sterilizer harness (5 and 6, fig. 6-7, sheet 1
(1) Position harness assembly (18, fig. 6-7, sheet 1
of 5).
of 5) in distribution box and install electrical leads.
(6) Remove the harness assemblies (5 and 6) from
(2) Replace washers(55, and 31, fig. 6-7, sheet2 of
the distribution box.
c. Disassembly. Disassemble the sterilizer harness
terminals of terminal boards (28, 38, 52 and 60) and
assemblies as required to replace damaged parts as
tighten nuts securely. Replace terminal board covers
(25, 35, 49 and 57).
(3) Place circuit breaker panel in position and secure
(2) Remove cable straps (2) and identification strap
with screws.
(3).
(3) Slide back insulation sleeving (4 and 5) to expose
6-23. Sterilizer Wiring Harness Assemblies
solder connections at connector (6) and unsolder
electrical wires (7) from connector (6). Remove sleeving
a.
General.
The sterilizer
wiring harness
(4 and 5).
assemblies consist of a connector and four electrical
leads to provide interconnection between the electrical
assemble the sterilizer harness assemblies as follows:
system and two sterilizers, when used.
b.
.Removal.
Remove
sterilizer harness
assemblies (5 and 6, fig. 6-7, sheet 1 of 5) according to
sequence of index numbers assigned to figure 6-7 and
1. Terminal (8, 4 per ay)
3. Strap, identification
5. Sleeving (6.3 per ay)
7. Wire
2. Strap, cable
4. Sleeving (2, 1 per ay)
6. Connector (2)
Figure 6-27. Sterilizer Wiring Harness Assembly (TS 013812)
6-37