TM 10-5410-221-14/1
(1) Place insulation sleeving (4 and 5) over wires (7)
terminal 5 on terminal board (28). Remove the electrical
and slide back to expose solder connections.
lead.
(2) Solder wires (7) to connectors in accordance with
(4) Remove screws and disconnect electrical leads
Specification MIL-S-6872 using Sn60 solder which
from circuit breaker (101, fig. 6-7, sheet 5 of 5).
conforms to Specification QQ-S-571.
(5) Remove nuts (7, fig. 6-7, sheet 1 of 5), washers
(3) Replace cable straps (2) and identification strap
(8) and screws (9) and remove connector (11) from the
(3).
distribution box.
(4) Install terminals (1) and crimp to secure.
(6) Remove the X-ray harness assembly (11) from
the distribution box.
c. Disassembly. Disassemble the X-ray harness
2), or other continuity checking device to check continuity
assembly (11, fig. 6-7, sheet 1 of 5) as required to
of wiring. Continuity must exist between end fittings on
individual wires. Continuity must not exist between
connector pins or between connector pins and connector
(2) Remove identification strap (3).
shell.
Replace any faulty wires or connectors as
(3) Slide back insulation sleeving (4 and 5) to expose
indicated during continuity test.
f. Repair. If possible, straighten any bent connector
soldered wire connectors to connector (6). Unsolder
connections and remove wires (7 and 8) from connector
pins. Inspect wires for abrasions and cuts of insulation.
(6). Remove sleeving (4 and 5).
Minor cuts and abrasions are acceptable and may be
d. Assembly. Assemble X-ray harness assembly
repaired by wrapping with vinyl tape. Replace individual
using figure 6-28 as a guide and observing the following:
wires if cut or otherwise damaged.
g.
Installation.
Install the sterilizer harness
(1) Place insulation sleeving (4 and 5, fig. 6-28) over
wires (7 and 8) and slide back to expose solder
assemblies, using figure 6-7 as a guide and observe the
connections.
following:
(2) Solder wires (7 and 8) to connector (6) in
(1) Position harness assemblies (5 and 6, fig. 6-7,
accordance with Specification MIIS-6872 using Sn60
sheet 1 of 5) in distribution box and install electrical
solder which conforms to Specification QQ-S-571.
leads.
(3) Replace identification strap (3).
(4) Install terminals (1 and 2) and crimp to secure.
(29) on terminals 4 and 5 of terminal board (38) and
tighten securely. Replace terminal board cover (35).
2), or other continuity checking device to check continuity
(3) Position connectors (6, fig. 6-27) in mounting
of wiring. Continuity must exist between end fittings on
holes and secure with screws (3, fig. 6-7, sheet 1 of 5),
individual wires. Continuity must not exist between
washers (2) and nuts (1).
connector pins or between connector pins and connector
(4) Position circuit breaker panel (123) on distribution
shell. Replace defective components, as required, if
box (124) and secure with screws.
indicated by continuity test.
f. Repair. If possible, straighten any bent connector
6-24. X-Ray Wiring Harness Assembly
pins. Inspect wires for abrasions and cuts of insulation.
Minor cuts and abrasions are acceptable and may be
a. General. The X-ray wiring harness assembly
repaired by wrapping with vinyl tape. Replace individual
consists of a connector and four electrical leads to
wires if cut or otherwise damaged.
provide interconnection between the electrical system
g. Installation. Install the X-ray harness using figure
and an X-ray unit.
6-7 as a
guide and observing the following:
Removal.
Remove X-ray wiring harness
b.
(1) Position harness assembly (11, fig. 6-7, sheet 1
assembly (11, fig. 6-7, sheet 1 of 5) according to
of 5) in distribution box and install electrical leads.
(2) Replace washers (31, fig. 6-7, sheet 2 of 5),
observing the following.
lockwasher (30) and nut (29) on terminal 5 of terminal
(1) Open circuit breaker panel cover.
board (28). Replace terminal board cover (25).
(2) Remove circuit breaker panel per paragraph 6-
21b.(2).
(3) Remove terminal board cover (25, fig. 6-7, sheet
2 of 5) and remove nut (29), lockwasher (30) and
6-38