TM 10-5410-221-24
1
Terminal (1)
3
Terminal (1)
5
Tie straps
2
Terminal (8)
4
Terminal (13)
6
Identification strap
Figure 6-23--Continued
electrical wires (7) from connector (6). Remove sleeving
25) in
mounting holes and secure with screws (3, fig. 6-
(4, 5).
22), washers (2) and nuts (1).
(4) Position circuit breaker panel (123) on
d. Assembly.
distribution box (124) and secure with screws.
Assemble sterilizer harness
assemblies in reverse order of disassembly using figure
6-25 as a
guide and observing the following.
6-26.
X-Ray Wiring Harness Assembly
(1) Place insulation sleeving (4, 5) over
wires (7) and slide back to expose solder connections.
a. General. The X-ray wiring harness assembly
(2) Solder wires (7) to connectors J5 and J6
consists of a connector and four electrical leads to
(6) in accordance with Specification MIL S-6872 using
provide interconnection between the electrical system
Sn60 solder which conforms to Specification QQ-S-571.
and an X-ray unit.
(3) Replace cable straps (2) and identification
b. Removal.
Remove X-ray wiring harness
strap (3).
assembly (11, fig. 6-22) according to sequence of index
(4) Install terminals (1) and crimp to secure.
following.
(1) Open circuit breaker panel cover.
continuity checking device on check continuity of wiring
(2) Remove circuit breaker panel per
individual wires. Continuity must not exist between
(3) Remove terminal board cover (25) and
connector pins or between connector pins and connector
remove nut (29), lockwasher (30) and terminal 5 on TB4
shell.
Replace any faulty wires or connectors as
terminal board (28). Remove the electrical lead.
indicated during continuity test.
(4) Remove screws and disconnect electrical
leads from CB3 circuit breaker.
f.
Repair.
If possible, straighten any bent
(5) Remove nuts (7), washers (8) and screws
connector pins. Inspect wires for abrasions and cuts of
(9) to remove J7 connector (11) from the distribution box.
insulation. Minor cuts and abrasions are acceptable and
(6) Remove the X-ray harness assembly (11)
may be repaired by wrapping with vinyl tape. Replace
from the distribution box.
individual wires if cut or otherwise damaged.
c. Disassembly. Disassemble the X-ray harness
g. Installation.
Install the sterilizer harness
assembly (11, fig. 6-22) as required to replace damaged
assemblies in reverse order of removal using figure 6-22
(1) Remove terminals (1, 2).
identification and observing the following.
(2) Remove identification strap (3).
(1) Position harness assemblies (5, 6, fig. 6-
(3) Slide back insulation sleeving (4, 5) to
22) in
distribution box and install electrical leads in
expose soldered wire connectors to connector (6).
Unsolder connections and remove wires (7, 8) from
connector (6). Remove sleeving (4, 5).
and nuts (29) on terminals 4 and 5 of TB3 terminal board
(38) and tighten securely. Replace terminal board cover
d. Assembly. Assemble X-ray harness assembly
(35).
in reverse order of disassembly using
(3) Position J5 and J6 connectors (6, fig. 6-
6-47